PERFORMANCE STUDY AND THERMAL PROFILE CHARACTERIZATION OF LOW TEMPERATURE VISCOSITY BATH IN DIFFERENT TEMPERATURE RANGES

H.M. Abu-Dorra; Mohy, E.; Mekkawy, M. M;

Abstract


A new Viscosity bath has been entered the services in Thermal Metrology Laboratory-National
Institute for Standards, NIS-Egypt in order to use in maintain and extend the national viscosity
scale in wide temperature ranges, international comparison and routine calibration of viscometers.
The medium of the bath should be homogenous enough in temperature so many thermal factors
taken into account to estimate the temperature gradient, homogeneity, stability and thermal
profile distribution with the related uncertainty to each parameter. The study carried out by two
Standards Platinum Resistance Thermometer (SPRT) calibrated at fixed point according to
ITS-90.


Other data

Title PERFORMANCE STUDY AND THERMAL PROFILE CHARACTERIZATION OF LOW TEMPERATURE VISCOSITY BATH IN DIFFERENT TEMPERATURE RANGES
Authors H.M. Abu-Dorra ; Mohy, E. ; Mekkawy, M. M 
Keywords Viscosity Bath, Viscometer, ITS-90, SPRT, Fixed Points, Uncertainty.
Issue Date Feb-2015
Publisher International Journal of Development Research
Journal International Journal of Development Research 

Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check

views 5 in Shams Scholar


Items in Ain Shams Scholar are protected by copyright, with all rights reserved, unless otherwise indicated.