Cyclic Fatigue Resistance of Three Different Rotary Ni-Ti Instruments

Marwa Mostafa Helmy Ghorab;

Abstract


The aim of the study was to evaluate the effect of autoclaving and curvature severity on Mpro,2Shape and Revo-s systems.
Total of ninty six files #25 with 6% of these brands were selected for the study, Cyclic fatigue testing was done using a cyclic fatigue testing device along with a custom made artificial stainless steel canals which had 60° and 3mm radius of curvature, and 45° and 5mm radius of curvature. There was three groups, each (n=32). Each group was subdivided into 2 subgroups according to sterilization: Subgroup a (non-sterilized) and Subgroup b (Sterilized for 5 cycles), each subgroup was further subdivided into two subdivisions according to the curvature and radius : Subdivision 1 (in the canal of 60° curvature and radius 3 mm.) and Subdivision 2 (in the canal of 45° curvature and radius 5 mm.) samples were rotated continuously until fracture in the artificial canals under body temperature to simulate clinical conditions.


Other data

Title Cyclic Fatigue Resistance of Three Different Rotary Ni-Ti Instruments
Other Titles مقارنة مقاومة المجهود الدوري بين ثلاثة أنظمة دوارة مختلفة من النيكل تيتانيوم
Authors Marwa Mostafa Helmy Ghorab
Issue Date 2019

Attached Files

File SizeFormat
CC294.pdf379.53 kBAdobe PDFView/Open
Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check

views 14 in Shams Scholar
downloads 9 in Shams Scholar