Spectrophotometric, Spectrofluorimetric and Potentiometric Assays of Cetyltrimethylammonium Bromide in Industrial Waste Water Samples"
Lobna Abdel Aziz Hussein; Nermine Victor Fares; Amira Mabrouk El-Kosasy;
Abstract
This work deals with spectrophotometric,
spectrofluorimetric, and potentiometric analyses of
cetyltrimethylammonium bromide (CTAB) cationic
detergent. The spectrophotometric procedure
depends on measuring the absorbance of its binary
complex with eosin yellow in Britton-Robinson
buffer (pH 4) at λmax 547 nm in the range of
2.0–14.0 μg/mL with an accuracy of 100.15 ± 0.54%.
The spectrofluorimetric procedure depends on
determining the quenching of the fluorescence
intensity of fluorescein dye by CTAB in the presence
of borate buffer at λem = 500 nm, λex = 304 nm, in the
range of 2.90–14.50 μg/mL with an accuracy of 99.81
± 0.33%. The electrochemical procedure describes an
ionophore-based technique using a graphite sensor
to measure 0.036 μg/mL and showed an accuracy
of 100.11 ± 0.61%. The experimental conditions
affecting each of the three suggested procedures
were studied and optimized. All the developed
procedures were validated and satisfactorily applied
for the determination of CTAB in industrial wastewater
samples.
spectrofluorimetric, and potentiometric analyses of
cetyltrimethylammonium bromide (CTAB) cationic
detergent. The spectrophotometric procedure
depends on measuring the absorbance of its binary
complex with eosin yellow in Britton-Robinson
buffer (pH 4) at λmax 547 nm in the range of
2.0–14.0 μg/mL with an accuracy of 100.15 ± 0.54%.
The spectrofluorimetric procedure depends on
determining the quenching of the fluorescence
intensity of fluorescein dye by CTAB in the presence
of borate buffer at λem = 500 nm, λex = 304 nm, in the
range of 2.90–14.50 μg/mL with an accuracy of 99.81
± 0.33%. The electrochemical procedure describes an
ionophore-based technique using a graphite sensor
to measure 0.036 μg/mL and showed an accuracy
of 100.11 ± 0.61%. The experimental conditions
affecting each of the three suggested procedures
were studied and optimized. All the developed
procedures were validated and satisfactorily applied
for the determination of CTAB in industrial wastewater
samples.
Other data
Title | Spectrophotometric, Spectrofluorimetric and Potentiometric Assays of Cetyltrimethylammonium Bromide in Industrial Waste Water Samples" | Authors | Lobna Abdel Aziz Hussein; Nermine Victor Fares ; Amira Mabrouk El-Kosasy | Issue Date | 2014 | Publisher | OXFORD UNIV PRESS INC | Journal | Journal of AOAC International | Volume | 97 | Issue | 4 | Start page | 1175 | End page | 1182 | DOI | 10.5740/jaoacint.13-102 |