EVALUATION OF SOLDERS FOR COBALT­ CHROMIUM PARTIAL DENTURES REPAIR

SHERINE MAMDOUH ABDEL HALIM;

Abstract


A cast model representing mandibular class I Kennedy's classification was obtained. The cast model was duplicated to obtain twelve refractory casts. For removable partial denture, Dentaurum alloy and Bego alloy were evaluated in an intact alloy and after soldering of the fractured framework at major and at minor connectors.


The flexural strengths for the two test alloys as intact and after soldering at major and minor connectors were measured and statistically analyzed. Some specimens from each group after fracture were selected and examined using Scanning electron microscope to evaluate the fracture surface.


The obtained results indicated that the intact major connector of Dentaurum alloy had the highest flexural strength followed by soldered major connector of Dentaurum alloy, intact major connector of Bego alloy, intact minor connector of Dentaurum alloy and least value in case of soldered minor connector of Bego alloy.


Other data

Title EVALUATION OF SOLDERS FOR COBALT­ CHROMIUM PARTIAL DENTURES REPAIR
Other Titles تقييم اللحام لتصليح الطقم الجزئى لسبيكة الكروم كوبالت
Authors SHERINE MAMDOUH ABDEL HALIM
Issue Date 2001

Attached Files

File SizeFormat
B13463.pdf957.96 kBAdobe PDFView/Open
Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check

views 4 in Shams Scholar


Items in Ain Shams Scholar are protected by copyright, with all rights reserved, unless otherwise indicated.