Temperature-Dependent Cyclic Fatigue Resistance and Differential Scanning Calorimetry Analysis of Three Rotary Nickel Titanium Instruments

Sara Sayed Bahr Idres;

Abstract


The aim of the study was to compare the CF resistance of Hyflex EDM, M3 Pro gold and Soco Pro files at room temperature and body temperature and to document the corresponding phase transformations.
Total of 60 files #25 of these brands were selected for the study, CF testing was done using a dynamic CF testing device along with a custom made artificial stainless steel canals which had 90° and 5mm radius of curvature. There was three groups, each (n=20). Each group was subdivided into 2 subgroups according to the environment temperature: Subgroup a (room temperature) and Subgroup b (body temperature), samples were rotated continuously until fracture in the artificial canals, the NCF was calculated. Phase transformation temperatures for 2 instruments of each brand were analyzed by DSC.


Other data

Title Temperature-Dependent Cyclic Fatigue Resistance and Differential Scanning Calorimetry Analysis of Three Rotary Nickel Titanium Instruments
Other Titles الإرهاق الدوري متأثرا بدرجة الحرارة وتحليل فرق المسح الكالوميتري لثلاثة أنظمة دوارة من النيكل تيتانيم
Authors Sara Sayed Bahr Idres
Issue Date 2020

Attached Files

File SizeFormat
BB1132.pdf763.79 kBAdobe PDFView/Open
Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check

views 7 in Shams Scholar


Items in Ain Shams Scholar are protected by copyright, with all rights reserved, unless otherwise indicated.