Fast neutron detection in CR-39 and DAM-ADC nuclear track detectors.

M. Eisa; A.M. Abdalla; O. Ashraf; Y.S. Rammah; A.H Ashry; T. Tsuruta;

Abstract


Fast detection of neutron in CR-39 and DAM-ADC nuclear track detectors were investigated using new etching conditions. The neutron irradiation is performed using a 5 mCi Am-Be source present at the National Institute of Standards (NIS) of Egypt. Using the new etching condition, irradiated CR-39 samples were etching for 4h and DAM-ADC samples for 80 min. Suitable analyzing software has been used to analyze experimental data. The dependence of neutron track density on the neutron fluence is investigated. When etching under optimum conditions, the relation between track density and fluence is determined which is found to be linear. Detection efficiency has been represented for both SSNTD, and found to be constant with fluence, which reflects the importance of using CR-39 and DAM-ADC detectors in the field of neutron dosimetry.
linear relationship between track density and effective dose is determined.


Other data

Title Fast neutron detection in CR-39 and DAM-ADC nuclear track detectors.
Authors M. Eisa ; A.M. Abdalla; O. Ashraf; Y.S. Rammah; A.H Ashry; T. Tsuruta
Keywords CR-39 | DAM-ADC | Neutron dosimetry | New etching conditions
Issue Date 11-Nov-2015
Publisher Radiation Physics and Chemistry
Journal Radiation Physics and Chemistry 
Volume 108
Start page 24
End page 28

Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check