Effect of dialyzer geometry on coagulation activation in the extracorporeal circuit in maintenance hemodialysis patients: Prospective randomized trial

ElSayed, Hesham; Sayed, Khalid Samir; Sary Gharib, Mohamed;

Abstract


Background and Objectives: The coagulation cascade is activated during hemodialysis (HD) due to interaction of blood with the dialysis circuit. There is a paucity of data on the effect of the physical structure of the dialyzers on coagulation activation during HD. We conducted this study to compare the effect of Helixone FX80 versus Platinum H4 dialyzers on coagulation activation during HD. Methods: Twenty patients on maintenance HD were enrolled in this randomized prospective crossover study. Each patient was dialyzed using Helixone FX80 and Platinum H4 dialyzers. Serum thrombin-antithrombin complex (TAT) was measured before (T0h) and at the end (T4H) of HD. Results: The absolute changes of serum TAT were comparable with the two dialyzers (median [IQR]: 1.15 [0.65, 1.75] for Helixone FX80 vs. 1.15 [0.67, 2.05] for Platinum H4, p = 0.371). Conclusion: Helixone FX80 and Platinum H4 dialyzers have similar effects on coagulation activation during HD.


Other data

Title Effect of dialyzer geometry on coagulation activation in the extracorporeal circuit in maintenance hemodialysis patients: Prospective randomized trial
Authors ElSayed, Hesham; Sayed, Khalid Samir; Sary Gharib, Mohamed 
Keywords coagulation activation;Helixone FX80;hemodialysis;Platinum H4;thrombin-antithrombin complex
Issue Date 1-Aug-2023
Journal Therapeutic Apheresis and Dialysis 
Volume 27
Issue 4
Start page 629
End page 635
ISSN 17449979
DOI 10.1111/1744-9987.13975
PubMed ID 36762467
Scopus ID 2-s2.0-85148427669

Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check

Citations 1 in scopus


Items in Ain Shams Scholar are protected by copyright, with all rights reserved, unless otherwise indicated.