Effect of Platform Switching on Strain Developed Around Implants Supporting Mandibular Overdenture

Sabet, Marwa E.; El-Mahrouky, Nesrin A.; El-Korashy, Dalia;

Abstract


Purpose: The aim of this study was to assess the effect of platform switching, using 3 different abutment sizes on the strain developed around implants supporting mandibular overdenture, using strain gages. MATERIALS: An implant-supported overdenture was constructed on an acrylic model with 2 implant fixtures placed at canine areas. Abutments of different sizes (5.5 mm, 4.5 mm, and 3.5 mm diameter) were screwed in the implant fixtures (platform switching). Anterior axial, anterior off-axis (lateral), and posterior load were applied on each abutment through a load applicator attached to the upper compartment of a universal testing machine until a load of 100 N, at which instant, the resultant strain on the implants was recorded using 4 strain gages connected to a 4-channel strain indicator. Data were analyzed by 1-way ANOVA at a significance level of 0.05, followed by Tukey-Kramer multiple comparisons test. RESULTS: Anterior axial load application and decreasing the abutment size lead to a decrease in the strain developed on the loaded abutment and to an increase in the strain developed on the unloaded abutment. Anterior off-axis load application and decreasing the abutment size lead to an increase in the strain developed on the loaded abutment and to a decrease on the unloaded abutment. On the other hand, posterior load application and decreasing the abutment size lead to an increase in the strain developed on the mesial side of the near abutment and a decrease in the strain developed on its distal side, where there was a significant increase in the strain developed on the mesial and distal sides of the far abutment. CONCLUSION: The distribution of the strain developed because of decreasing the abutment size does not favor the use of platform switching in an implant-supported mandibular overdenture. © 2009 Lippincott Williams & Wilkins, Inc.


Other data

Title Effect of Platform Switching on Strain Developed Around Implants Supporting Mandibular Overdenture
Authors Sabet, Marwa E.; El-Mahrouky, Nesrin A.; El-Korashy, Dalia 
Keywords Abutment size;Implant;Overdenture;Platform switching;Strain gages
Issue Date 1-Aug-2009
Journal Implant Dentistry 
Volume 18
Issue 4
Start page 362
End page 370
ISSN 10566163
DOI 10.1097/ID.0b013e3181b4f98c
PubMed ID 19667825
Scopus ID 2-s2.0-69849093235

Recommend this item

Similar Items from Core Recommender Database

Google ScholarTM

Check



Items in Ain Shams Scholar are protected by copyright, with all rights reserved, unless otherwise indicated.